SYNTHES GMBH T-HANDLE BALL HEX SCREWDRIVER 8MM Back to Search Results
Catalog Number 03.010.517
Medical Device Problem Codes Break (1069); Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in netherlands as follows: it was reported that on march 20, 2024, during a tfna procedure the device broke.The doctor had to change the tfna implant because the insertion handle could not be removed.A new sent was obtained and the implant was changed.There was a forty-five (45) minute surgical delay.This report involves one (1) t-handle ball hex screwdriver 8mm.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d4: lot.H3, h4 h6: photos were provided for review.The photo investigation revealed that '03.010.517, scrdriver-hex 8 w/t-handle w/spheric-he had broken tip.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.For the other condition functional issue cannot determined by photo.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Correction: g1: manufacturing site name and address.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: the product was returned to depuy synthes for evaluation.Visual inspection of the returned device revealed that the tip of the scrdriver-hex ø8 w/t-handle w/spheric-he was found broken.The broken fragment was returned for examination.It is reasonable to conclude that this broken condition may prevent the assembly with the mating device.Therefore, the reported condition can be confirmed.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.A dimensional inspection was unable to be performed due to post manufacturing damage.A functional evaluation was unable to be performed due to post manufacturing damage.The overall complaint was confirmed as the observed condition of the scrdriver-hex ø8 w/t-handle w/spheric-he would contribute to the complained device issue.¿ based on the investigation findings, the potential cause is traced to component failure and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): product code: : 03.010.517 lot number : 9945400 release to warehouse date : 20.Jul.2016 expiration date : na supplier: na manufacturing site: werk hagendorf.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-HANDLE BALL HEX SCREWDRIVER 8MM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19053470
Report Number8030965-2024-04596
Device Sequence Number12530095
Product Code HXX
UDI-Device Identifier07611819499133
UDI-Public07611819499133
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.010.517
Device Lot Number9945400
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer04/23/2024
05/29/2024
Initial Report FDA Received Date04/05/2024
Supplement Report FDA Received Date05/21/2024
06/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/20/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - NAILS: TFNA.
Patient SexUnknown
-
-