| Catalog Number |
23533103 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Unequal Limb Length (4534)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the surgery was performed on (b)(6), but the two proximal lag screws were inserted off the nail.The proximal bone fragment slid 7-8 cm distally, and the femur has shortened.Revision surgery was performed on (b)(6).".
|
| |
|
Event or Problem Description
|
|
As reported: "the surgery was performed on (b)(6), but the two proximal lag screws were inserted off the nail.The proximal bone fragment slid 7-8 cm distally, and the femur has shortened.Revision surgery was performed on (b)(6).".
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: d1 and h6 method code.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.In case substantive information will become available in future that suggests otherwise we reserve the right to reopen the case.
|
| |
|