• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES TRAUMA : CROSSROADS EXTREMITY SYSTEMS, LLC POLYAXIAL LCKNG SCREW 3.5MM X 18MM STER; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES TRAUMA : CROSSROADS EXTREMITY SYSTEMS, LLC POLYAXIAL LCKNG SCREW 3.5MM X 18MM STER; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 15PL-3518
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a revision surgery occurred on (b)(6) 2024 due to pain and the migration of the screws backing out of the dynabunion plate.The two proximal screws were explanted via removal procedure.Original implant date was (b)(6) 2023.Patient was healed at the time of revision, joint fusion had occurred per the surgeon, and successful removal of painful hardware completed.The procedure was completed successfully.This report is for one motobnd cp nonlkng scr 3.5mm x 18mm ster for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Unknown when migration occurred.E1: initial reporter is j&j company representative without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Corrected data: d4.Additional narrative: h6: a manufacturing record evaluation cannot be performed due to lot number being unknown.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo and x-ray revealed that the polyaxial lckng screw 3.5mm x 18mm ster exhibits signs of nicked in the threaded area and wear on the surface consistent with implantation and removal process.The defect of migration can be confirmed as it can be observed that two screws are protruding from the plate and are not completely fixed to it.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for polyaxial lckng screw 3.5mm x 18mm ster there is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL LCKNG SCREW 3.5MM X 18MM STER
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES TRAUMA : CROSSROADS EXTREMITY SYSTEMS, LLC
6423 shelby view dr.
suite 101
memphis TN 38134
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19055690
Report Number3020584246-2024-00007
Device Sequence Number7339156
Product Code HWC
UDI-Device Identifier00815432026807
UDI-Public(01)00815432026807
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K160300
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number15PL-3518
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/21/2024
Supplement Date Received by Manufacturer05/20/2024
Initial Report FDA Received Date04/05/2024
Supplement Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MOTOBND CP NONLKNG SCR 3.5MM X 18MM STER; POLYAXIAL LCKNG SCREW 3.5MM X 18MM STER; UNK - PLATES: TRAUMA
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-