ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM Y-TYPE BLOOD SET, 200 MICRON FILTER, CLAVE SECONDARY PORT, CLAVE Y-; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
14220-01 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received for investigation.Investigation is pending.
|
| |
|
Event or Problem Description
|
|
The event occurred on an unspecified date involving a primary plum y-type blood set, 200 micron filter, clave secondary port, clave y-site, non-vented, secure lock, 110 inch where the customer reported to have cases that have holes in it.There was unknown patient involvement and unknown patient harm.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Received one-hundred forty-four (144) new list #14220-01 primary plum y-type blood sets.As received small evenly spaced holes were observed on the exterior packaging of the sets.Those small holes are purposely put in to allow for the air to escape.Packaging icon denotes these are sterile fluid path sets.The caps on the ends of the sets maintain sterility of the internals of the sets.The complaint of pin hole in plastic wrapping can be confirmed.However, these small holes are intentional and do not affect the form, fit, function or sterility of the product.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
| |
|