| Catalog Number |
723-10-32F |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Joint Dislocation (2374)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported by sales representative that patient did rthr revision on (b)(6) 2024.Had his right hip dislocated and proximal femur fracture in early february when patient crossed his legs.Been on traction treatment, and did his second rthr revision.Dr.Changed from normal trident 2 liner to constrained liner.Also, dall miles beaded wire was sued to fix the proximal femur fracture.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding dislocation involving a trident liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant indicated: "confirmation of event: i can confirm that the patient had a right total hip arthroplasty at some point in the past with a cementless cup and a long stem cementless femoral component.I can also confirm that the patient developed a fracture of the greater trochanter.I was able to see x-rays with the implant and the fracture in place.Root cause analysis: the root cause of this event cannot be determined with certainty.The causes of late fracture of the greater trochanter following total hip arthroplasty are multifactorial including surgical technique and patient factors including lifestyle, activity level and bmi.The cause of the initial fracture cannot be determined with certainty but trauma may have played a role.After revision the patient apparently crossed his legs and this apparently contributed to a re-fracture after the first revision.If fixation was not adequate than crossing one's leg in the initial postoperative period could cause displacement again.I would not assign any causality to the implant itself." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to dislocation and periprosthetic fracture of the proximal femur.A review of the medical records by a clinical consultant was unable to confirm the dislocation event: "i can confirm that the patient had a right total hip arthroplasty at some point in the past with a cementless cup and a long stem cementless femoral component.I can also confirm that the patient developed a fracture of the greater trochanter.I was able to see x-rays with the implant and the fracture in place.Root cause analysis: the root cause of this event cannot be determined with certainty.The causes of late fracture of the greater trochanter following total hip arthroplasty are multifactorial including surgical technique and patient factors including lifestyle, activity level and bmi.The cause of the initial fracture cannot be determined with certainty but trauma may have played a role.After revision the patient apparently crossed his legs and this apparently contributed to a re-fracture after the first revision.If fixation was not adequate than crossing one's leg in the initial postoperative period could cause displacement again.I would not assign any causality to the implant itself." no further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported by sales representative that patient did rthr revision on (b)(6) 2024.Had his right hip dislocated and proximal femur fracture in early february when patient crossed his legs.Been on traction treatment, and did his second rthr revision.Dr.Changed from normal trident 2 liner to constrained liner.Also, dall miles beaded wire was sued to fix the proximal femur fracture.
|
| |
|