| Model Number |
FR4A-TRL-B0 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The commercial team member reported that the implanting clinician was unable to safely advance the introducer to implant the trial neurostimulator without moving/dragging surrounding tissue with it.The trial procedure was aborted.The patient received a nerve block and will search for other pain relief options.
|
| |
|
Additional Manufacturer Narrative
|
|
The other adverse events questionnaire was reviewed for potential causes of the reported issue.The implanting clinician stated that the reported issue was due to patient anatomy as the patient's skin was too loose which prevented the introducer from advancing.The stimulator is used to treat pain.The cause of the reported issue is due to the patient being a poor candidate due to health, weight, age, mental capacity as the patient's skin was too loose which prevented the introducer from advancing (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in other adverse events issues.Other adverse events issues rates remain acceptably low; thus, a capa is not required.Other adverse events issues rates will continue to be tracked and trended.
|
| |
|
Event or Problem Description
|
|
The commercial team member reported that the implanting clinician was unable to safely advance the introducer to implant the trial neurostimulator without moving/dragging surrounding tissue with it.The trial procedure was aborted.The patient received a nerve block and will search for other pain relief options.
|
| |
|
Additional Manufacturer Narrative
|
|
The other adverse events questionnaire was reviewed for potential causes of the reported issue.The implanting clinician stated that the reported issue was due to patient anatomy as the patient's skin was too loose which prevented the introducer from advancing.The stimulator is used to treat pain.The cause of the reported issue is due to the patient being a poor candidate due to health, weight, age, mental capacity as the patient's skin was too loose which prevented the introducer from advancing (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in other adverse events issues.Other adverse events issues rates remain acceptably low; thus, a capa is not required.Other adverse events issues rates will continue to be tracked and trended.
|
| |
|