• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE 2.0/2.4MM CANNULATED SCREW DRIVER BIT; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE 2.0/2.4MM CANNULATED SCREW DRIVER BIT; SCREWDRIVER Back to Search Results
Catalog Number 210-24-003
Medical Device Problem Codes Mechanical Problem (1384); Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported by independent sales representative, the driver broke when the dr had the screw halfway into bone.She heard a crack and metal broke off the driver and was unable to capture the screw head.There was a delay in the case to try and find a solution.The explanted screws were discarded.Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Instrument failure - 20 minute delay in surgery.
 
Additional Manufacturer Narrative
One similar complaints was identified for a tiger headed driver breaking.The root cause of the similar complaints is unknown due to limited information.There were no abnormalities in regards to dhr review.It is unknown if the doctor followed the surgical technique with the limited information available.The broken driver was not returned.Thus, full visual inspection and dimensional inspection did not occur.Limited visual inspection occurred utilized the reporter provided photograph to confirm the damage and the lot number.Simulated use testing did not occur as the driver was damaged during a hardware removal.Based on the limited information available regarding technique, as well as the devices not being retuned, the root cause shall remain as unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0/2.4MM CANNULATED SCREW DRIVER BIT
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE
727 north shepherd drive
suite 100
houston TX 77007
Manufacturer (Section G)
TRILLIANT SURGICAL, LLCD/BA DJO FOOT AND ANKLE
727 north shepherd drive
suite 100
houston TX 77007
Manufacturer Contact
stacey bonnell
727 north shepherd drive
suite 100
houston, TX 77007
8777883864
MDR Report Key19056509
Report Number3007420745-2024-00001
Device Sequence Number6677064
Product Code HXX
UDI-Device Identifier00812926021393
UDI-Public00812926021393
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
D300093
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number210-24-003
Device Lot NumberTSL016232
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2024
Supplement Date Received by Manufacturer05/08/2025
Initial Report FDA Received Date04/06/2024
Supplement Report FDA Received Date07/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age51 YR
Patient SexFemale
-
-