| Model Number |
BELLAVISTA 1000E US 17,3 VENTILATOR |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H3:81 other: the suspect device has not been return for investigation.Furthermore, no root cause has been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.H3 other text : device not returned yet.
|
| |
|
Event or Problem Description
|
|
It was reported to vyaire medical that the bellavista1000e us 17,3" ventilator had an error message: "bellavista detected a user interface issue - decommission bv" screen.Furthermore, patient was transferred to alternative ventilation and no harm associated with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: vyaire medical field service representative went onsite to check and troubleshoot the device.The root cause was determined due to software issue.Most likely it is caused by ui overload.Investigations performed on similar cases by imt proved that this failure can be reproduced in the lab and the ventilation keep continuing with the last applied setting.This failure classified as "permanent apphang while device in standby mode".As a resolution, software bug fix improvements will be implemented on next sw release.Vyaire medical fsr ran verification per instruction from support.All tests passed required criteria.Unit is fully operational & available to service.
|
| |
|