Catalog Number 47177E
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd alaris smartsite gravity set was damaged the following information was received by the initial reporter with the following verbatim: one of our intermountain facilities is reporting an issue/concern with an item that it is purchased from your company.Situation: issue: the tubing used for a stem cell boost snapped right below the chamber during the infusion.
 
Additional Manufacturer Narrative
It was reported that the tubing snapped just below the drip chamber.One sample model 47177e, lot 23079289 was returned for investigation.The set was examined for defects and abnormalities.The tubing was separated from the drip chamber.No other defects or abnormalities were observed.The customer complaint was verified.From the manufacturer's investigation, the potential root cause could be related to failures of the solvent dispenser and assembly process does not have support that ensures the correct dimension and alignment of the tubing to ensure solvent is dispensed on the tubing.H3 other text : see h10 below.
 
Event or Problem Description
No additional info.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE GRAVITY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key19056983
Report Number9616066-2024-00540
Device Sequence Number20665306
Product Code FPA
UDI-Device Identifier10885403235436
UDI-Public(01)10885403235436
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date07/26/2026
Device Catalogue Number47177E
Device Lot Number23079289
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2024
Supplement Date Received by Manufacturer05/03/2024
Initial Report FDA Received Date04/07/2024
Supplement Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1