Catalog Number DYNJ50788F
Medical Device Problem Codes Suction Problem (2170); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a bulb irrigation syringe component broke "apart upon use.".
 
Additional Manufacturer Narrative
It was reported that a bulb irrigation syringe component broke "apart upon use." no serious injury or adverse impact to a patient or a user was originally reported.To date, no information has been received to indicate that a user or a patient experienced a death, serious injury, medical intervention, follow-up care, or other adverse health impact associated with the reported problem/issue.A sample was returned for evaluation and testing identified a lack of suction and a small base diameter which would not allow the bulb to properly connect to the syringe barrel.In an abundance of caution, and in response to an fda 483 issued for cfn (b)(4) on 22-jan-2024, this medwatch is being filed for the reported problem/issue.If additional relevant information becomes available a supplemental medwatch will be filed.
 
Additional Manufacturer Narrative
Updated to include expiration date of device which was not included in original report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
BREAST PACK
Manufacturer (Section D)
MEDLINE INDUSTRIES, LP
three lakes drive
northfield IL 60093 2753
Manufacturer Contact
kelly zampella
three lakes drive
northfield, IL 60093-2753
MDR Report Key19057220
Report Number1423395-2024-00323
Device Sequence Number6593301
Product Code FTN
UDI-Device Identifier40195327421039
UDI-Public40195327421039
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/30/2025
Device Catalogue NumberDYNJ50788F
Device Lot Number23HBI983
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/11/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date10/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-