• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTONIX, INC LUTONIX 035AV DRUG COATED PTA DILATION CATHETER; DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTONIX, INC LUTONIX 035AV DRUG COATED PTA DILATION CATHETER; DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Restenosis (4576); Insufficient Information (4580)
Date of Event 02/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of a clinical trial that sometime post an index procedure using a drug-coated balloon catheter, the subject reportedly expired and the primary cause of death was not provided.Reportedly, there have been no documented device deficiencies and no adverse events were reported, and the relationship of the death with the device, procedure and av access circuit were not provided.
 
Additional Manufacturer Narrative
H10: additional information was received, and the cause of death was not related to the device, not related to the procedure and not related to the av access circuit.Therefore, the file was reassessed for reportability and determined to be serious injury.H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.D4 (expiry date: 04/2025).H11: b2, b3, b5, d1, d4, h1, h6 (patient).The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the results of a clinical trial that six months and seven days post an index procedure completion using a drug-coated balloon catheter, the subject had an adverse event of restenosis of the target lesion and required an arteriovenous access circuit reintervention.It was further reported that the target lesion was in the upper arm with eighty percent initial stenosis and zero percent final residual stenosis.Reportedly, standard percutaneous transluminal angioplasty was used for treatment.The re-intervention was successful, and the outcome of the adverse event was resolved.It was further reported that patient was expired and the event was not related to device, procedure and av access circuit.
 
Event or Problem Description
It was reported through the results of a clinical trial that six months and seven days post an index procedure completion using a drug-coated balloon catheter, the subject had an adverse event of prolonged bleeding and restenosis of the target lesion and required an arteriovenous access circuit reintervention.It was further reported that the target lesion was in the upper arm with eighty percent initial stenosis and zero percent final residual stenosis.Reportedly, standard percutaneous transluminal angioplasty was used for treatment.The re-intervention was successful, and the outcome of the adverse event was resolved.It was further reported that patient was expired and the event was not related to device, procedure and av access circuit.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported event as no objective evidence was provided for review.Based upon the available information, the definitive root cause for the reported event could not be determined.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, g3, h6 (patient, method).H11: d4 (unique identifier (udi) #).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 035AV DRUG COATED PTA DILATION CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope MN 55428
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19057472
Report Number3006513822-2024-00016
Device Sequence Number3568946
Product Code PRC
UDI-Device Identifier00801741135439
UDI-Public(01)00801741135439(11)220603(17)250425(10)GFGQ2725
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P170003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/09/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2025
Device Model Number9010
Device Catalogue NumberLX3575760V
Device Lot NumberGFGQ2725
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer06/11/2024
09/19/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date07/09/2024
10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Death;
Patient Age37 YR
Patient SexFemale
Patient Weight50 KG
Patient EthnicityHispanic
Patient RaceWhite
-
-