BIOMET UK LTD. OXFORD ANAT BRG RT MD SIZE 3 PMA; KNEE PROSTHESIS Back to Search Results
Catalog Number 159575
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374)
Date of Event 03/07/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10 associated medical devices: oxf uni tib tray sz c rm pma; item# 154723; lot# 3819016 oxford uni twin-peg femoral md; item# 166942; lot# j3670783 g2 - foreign: japan the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event or Problem Description
It was reported that post an initial uka surgery, the patient experienced a bearing dislocation due to the the fracture of said bearing which resulted in revision surgery.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h2, h3, h6, h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that could be related to the reported event.Radiographs were provided and reviewed by a radiologist.The review identified : a medial unicompartmental arthroplasty has been performed.There is no evidence of implant loosening.The radiopaque markers of the bearing are noted superolaterally consistent with dislocation of the bearing, likely into the lateral suprapatellar recess.There is narrowing of the medial implant tibiofemoral space and early lateral compartment arthrosis.Implant fit and alignment are overall maintained despite the bearing dislocation.Bone quality is osteopenic.The bearing dislocation is confirmed on the radiograph, but a bearing fracture cannot be confirmed on this image.Other than the bearing dislocation, no abnormality is identified.With the available information, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD ANAT BRG RT MD SIZE 3 PMA
Common Device Name
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key19058278
Report Number3002806535-2024-00128
Device Sequence Number6753003
Product Code NRA
UDI-Device Identifier05019279786213
UDI-Public(01)05019279786213(17)210520(10)3794388
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/20/2021
Device Catalogue Number159575
Device Lot Number3794388
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2024
Supplement Date Received by Manufacturer05/02/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date05/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
-
-