• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRECISION SPINE, INC VAULT C ACDF SYSTEM; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRECISION SPINE, INC VAULT C ACDF SYSTEM; SCREWDRIVER Back to Search Results
Catalog Number 37-IN-0060
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a procedure was performed in the dominican republic on (b)(6) 2024, utilizing the vault c acdf system.After preparing the disc space, they used a rasp to refine the endplate.Following that, they inserted the implant into the disc and prepared the screw hole.They utilized a straight awl and then a straight drill.While attempting to insert a screw, the tip of the self-retaining driver (37-in-0060) broke.The tip was unable to be removed from the head of the screw so the screw was removed and replaced utilizing the angled driver readily available in the set.There was a one (1) minute delay to the procedure due to the malfunction.No patient injury was reported.
 
Additional Manufacturer Narrative
Evaluation in process but not yet complete.Upon completion of evaluation, a follow up report will be submitted.
 
Additional Manufacturer Narrative
H3 device evaluation - engineering performed a visual inspection of the complaint product.The distal tip of the t8 self-retaining driver, (b)(6), is completely fractured in a plane that is perpendicular to the shaft axis.The cause for the fracture is due to excessive torque experienced during this procedure or due to this and prior high loading conditions that may have results in cracks and manifested itself in this failure.The device history record confirms that the 40 piece lot had been inspected and conform to drawing specifications.Two year complaint history review did not reveal a trend for reports of this nature for this part number.The need for corrective action is not indicated at this time however, complaints will continue to be monitored.
 
Event or Problem Description
It was reported that a procedure was performed in the dominican republic on (b)(6) 2024, utilizing the vault c acdf system.After preparing the disc space, they used a rasp to refine the endplate.Following that, they inserted the implant into the disc and prepared the screw hole.They utilized a straight awl and then a straight drill.While attempting to insert a screw, the tip of the self-retaining driver (b)(6) broke.The tip was unable to be removed from the head of the screw so the screw was removed and replaced utilizing the angled driver readily available in the set.There was a one (1) minute delay to the procedure due to the malfunction.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAULT C ACDF SYSTEM
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer (Section G)
PRECISION SPINE, INC
2050 executive drive
pearl MS 39208
Manufacturer Contact
mike dawson
2050 executive drive
pearl, MS 39208
6014204244
MDR Report Key19060840
Report Number3005739886-2024-00022
Device Sequence Number6752605
Product Code HXX
UDI-Device Identifier00840019916974
UDI-Public00840019916974
Combination Product (Y/N)N
Initial Reporter CountryDR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number37-IN-0060
Device Lot Number49059MI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer03/13/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date04/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-