| Model Number |
328418 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
2 syringes affected by this event.H3 other text: device not available.
|
| |
|
Event or Problem Description
|
|
Consumer reported while drawing from her ozembec vial, the needle broke.Denied reuse of needles.Happened with 2 different syringes today.Lot #: 3219478.Catalog#: 328418.Date of event: 03.13.2024.Sample status: discard.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to medwatch section c for the affected product is attached.Correction to: (type of investigation, and investigation conclusions) investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|