| Model Number |
BREV100 |
| Medical Device Problem Code |
Failure to Obtain Sample (2533)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.A field engineer examined the device and found that there was an error in the system.The field engineer cleared the error and the unit functioned normally.Check unit for functionality and found no issues.Possibly the driver was not loaded into the unit with the gears meshing.The unit performed to manufacture standard.If additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) a brevera procedure was performed, and during the procedure, a driver error occurred, it took two punctures to the skin and 3 needles to complete the procedure.No additional information available.
|
| |
|