• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V+C, JAPANESE
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2024
Type of Reportable Event Malfunction
Event or Problem Description
An authorized third party service provider (asp) contacted ventec to report that the ventilator measured lower peep (positive end expiratory pressure) than set.There were no reports of patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
H6: the device was not returned to ventec for evaluation.The device was further evaluated by the authorized service provider (asp) where the reported issue of it measuring lower peep (positive end expiratory pressure) than set was confirmed.The asp replaced the exhalation control module (ecm) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the ecm.H3 other text : serviced by asp.
 
Additional Manufacturer Narrative
Corrected information for supplemental medwatch report 001 -- section d4, model #, of the initial medwatch report stated: prt-01098-100, section d4, model #, of the initial medwatch report should have stated: v+c, japanese, section d4, udi #, of the initial medwatch report stated: (b)(4), section d4, udi #, of the initial medwatch report should have stated: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key19066815
Report Number3013095415-2024-00307
Device Sequence Number20341022
Product Code CBK
UDI-Device Identifier00850018761116
UDI-Public(01)00850018761116(11)210225(21)122999
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation 505
Type of Report Initial,Followup
Report Date (Section B) 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV+C, JAPANESE
Device Catalogue NumberPRT-01098-100
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2024
Supplement Date Received by Manufacturer03/18/2026
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-