| Model Number |
D021 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/22/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) had an alert indicating that there was an out of range pacing impedance value on the right ventricular (rv) lead channel.Technical services (ts) reviewed the reports and advised monitoring the issue as the impedance values were slightly below the nominal upper limit.Ts noted there were no noisy signals in the logbook or in the presenting.The device remains in use.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) had an alert indicating that there was an out of range pacing impedance value on the right ventricular (rv) lead channel.Technical services (ts) reviewed the reports and advised monitoring the issue as the impedance values were slightly below the nominal upper limit.Ts noted there were no noisy signals in the logbook or in the presenting.The device remains in use.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.Investigation has determined that this type of event is likely the result of an intermittent high impedance condition associated with the device spring contact and lead terminal ring.A design enhancement was implemented in 2020 to stabilize the electrical connection between the spring contact and the lead terminal.
|
| |
|
Search Alerts/Recalls
|