| Model Number |
3.5MM PUSHLOCK, BIOCOMP, SELF-PUNCHING |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
On 3/18/2024, it was reported by a sales representative via email that an ar-1926bcsp biocomposite pushlock anchor eyelet broke on impaction.The broken fragments were removed from inside the patient.This was discovered during an rcr procedure on (b)(6)2024.Additional information received on 3/26/2024: there was no delay in the procedure, and additional anesthesia was not needed.The patient suffered no negative effects.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint allegation is confirmed.Visual evaluation of the attached picture revealed that the eyelet implant had fractured at the junction between the base of the eyelet and the threaded tip of the shaft, which is the point of connection to the driver.Functional testing was not performed due to the device's damage.Bone quality and bone preparation were provided.The most likely causes of reported failure are misaligned insertion or prying/leveraging the device during insertion.
|
| |
|