• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH WIRE TENSIONER HOFFMANN LRF; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH WIRE TENSIONER HOFFMANN LRF; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 49339100
Medical Device Problem Codes Break (1069); Fracture (1260); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/15/2023
Type of Reportable Event Malfunction
Event or Problem Description
As reported: " during the cleaning and sterilization , instrument has been broken.Do not know what happened, but parts do not fit like they have done before.They have used cleaning protocol advice we have in writing (ifu).No harm in operation when used, only in cleaning process have noticed broken device.No patient harm or delay.".
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
Corrections: please refer to section h6 (results code).The complaint could be confirmed, since the product was returned for evaluation and matches the alleged failure.The device inspection revealed the following: visual examination of the received product shows that the received product only have some dents and scratches at the black handle.Otherwise there are no visible damage are shown.Mechanical engineering reviewed the received information and noted: during the investigation could be determined that the adjusting ring with which the tensile force is set and locked with "loctite" is loose.This leads to the fact that the setted, calibrated, and tested force no longer corresponds to the markings.In a further investigation, the adjustment ring was loosened, it was found that no adhesive residue (loctite) was visible.This suggests that the ring was probably assembled incorrectly with less or without loctite.Inspection of the received information/evidences are done or will be done, in the proceedings of the nc.If more information is provided, the case will be reassessed.
 
Event or Problem Description
As reported: " during the cleaning and sterilization , instrument has been broken.Do not know what happened, but parts do not fit like they have done before.They have used cleaning protocol advice we have in writing (ifu).No harm in operation when used, only in cleaning process have noticed broken device.No patient harm or delay.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WIRE TENSIONER HOFFMANN LRF
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19067470
Report Number0008031020-2024-00152
Device Sequence Number17609920
Product Code LXH
UDI-Device Identifier07613252611035
UDI-Public07613252611035
Combination Product (Y/N)N
Initial Reporter CountryFI
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number49339100
Device Lot NumberJ43995
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/05/2024
Supplement Date Received by Manufacturer06/17/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date07/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-