| Catalog Number |
49339100 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported: " during the cleaning and sterilization , instrument has been broken.Do not know what happened, but parts do not fit like they have done before.They have used cleaning protocol advice we have in writing (ifu).No harm in operation when used, only in cleaning process have noticed broken device.No patient harm or delay.".
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrections: please refer to section h6 (results code).The complaint could be confirmed, since the product was returned for evaluation and matches the alleged failure.The device inspection revealed the following: visual examination of the received product shows that the received product only have some dents and scratches at the black handle.Otherwise there are no visible damage are shown.Mechanical engineering reviewed the received information and noted: during the investigation could be determined that the adjusting ring with which the tensile force is set and locked with "loctite" is loose.This leads to the fact that the setted, calibrated, and tested force no longer corresponds to the markings.In a further investigation, the adjustment ring was loosened, it was found that no adhesive residue (loctite) was visible.This suggests that the ring was probably assembled incorrectly with less or without loctite.Inspection of the received information/evidences are done or will be done, in the proceedings of the nc.If more information is provided, the case will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: " during the cleaning and sterilization , instrument has been broken.Do not know what happened, but parts do not fit like they have done before.They have used cleaning protocol advice we have in writing (ifu).No harm in operation when used, only in cleaning process have noticed broken device.No patient harm or delay.".
|
| |
|