• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 SYRINGE 1.0ML 27GA 5/8IN 10BAG 500 BDD; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949 SYRINGE 1.0ML 27GA 5/8IN 10BAG 500 BDD; SYRINGE, PISTON Back to Search Results
Model Number 329412
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3 other text : device is not available.
 
Event or Problem Description
A pharmacy technician was dispensing product (b)(6) - lot 309 3588, caused a needle stick due to either the cap missing or the cap malfunctioning.The needle stick was a clean stick and the technician checked the rest of the needles and no others were missing the caps.
 
Additional Manufacturer Narrative
Additional information provided to sections.Corrections made to section h6 (component code, type of investigation, & investigation conclusions).Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 1.0ML 27GA 5/8IN 10BAG 500 BDD
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE - HOLDREGE, NE / 68949
Manufacturer (Section G)
BECTON, DICKINSON & CO.
1 becton dr.
franklin lakes NJ 07417
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key19067900
Report Number1920898-2024-05077
Device Sequence Number12229964
Product Code FMF
UDI-Device Identifier30382903294122
UDI-Public30382903294122
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K190054
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number329412
Device Lot Number3093588
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date07/03/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-