MAUDE Adverse Event Report: ABBOTT VASCULAR 20/30 PRIORITY PACK ACCESSORY KIT; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR 20/30 PRIORITY PACK ACCESSORY KIT; ACCESSORIES Back to Search Results
Catalog Number 1000186
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/19/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.The device history record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other incidents from this lot.It is possible that the connection port was not fully connected/tightened and/or the device being connected was compromised thus resulting in the reported leak difficulties; however as the device was not returned for analysis this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported leak difficulties.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that during a procedure was to treat the right coronary artery.The indeflator failed to inflate the balloon when applying pressure.Another indeflator was used to successfully complete the procedure.There was no adverse patient effects reported and no clinically significant delay.No additional information was provided.
 
Additional Manufacturer Narrative
H6: health effect - impact code 4648 removed and 2199 added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20/30 PRIORITY PACK ACCESSORY KIT
Common Device Name
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key19068044
Report Number2024168-2024-04437
Device Sequence Number4732495
Product Code MAV
UDI-Device Identifier08717648013614
UDI-Public(01)08717648013614(17)250930(10)60511562
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K961471
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Catalogue Number1000186
Device Lot Number60511562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/21/2024
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date04/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-