| Catalog Number |
1000186 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.The device history record and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other incidents from this lot.It is possible that the connection port was not fully connected/tightened and/or the device being connected was compromised thus resulting in the reported leak difficulties; however as the device was not returned for analysis this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported leak difficulties.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure was to treat the right coronary artery.The indeflator failed to inflate the balloon when applying pressure.Another indeflator was used to successfully complete the procedure.There was no adverse patient effects reported and no clinically significant delay.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: health effect - impact code 4648 removed and 2199 added.
|
| |
|