• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO12.6
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Toxic Anterior Segment Syndrome (TASS) (4469)
Date of Event 03/09/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A4-a6:unk.H6: work order search: one similar complaint within associated lots were found.Claim# (b)(4), fellow right eye (od).Manufacturer narrative: a review of the device labeling was completed.Anterior chamber empyema, fibrin precipitation, and iritis are identified in the labeling as known potential adverse events following icl implantation.The dfu provides the surgeon instruction for complete ovd removal.Precaution: (6) after inserting this product, aspirate the viscoelastic substance completely from inside the eye.Do not use highly viscous viscoelastic substances that are difficult to aspirate completely.Toxic anterior segment syndrome (tass) is an acute sterile postoperative inflammation that can occur after uncomplicated or complicated intraocular surgery.While improper sterilization and contamination of surgical instruments remains the most common risk factor associated with tass, ocular viscoelastic, and improper preparation of antibiotics for intracameral injection may also lead to tass.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that a 12.6mm vicmo -15.00d implantable collamer lens was implanted into the patient's right eye (os) on (b)(6) 2024.Same day bilateral icl implantation was performed.Concomitant products used intraoperatively include opegan viscoelastic and the msi injector system.On day 2, the patient experienced significant vision loss and went to hospital where anterior chamber empyema (hypopyon) with a fibrin membrane was noted in the pupillary area.No cultures were taken.Treatment with rinderon subconjunctival injection and sunbeta hourly were given.After anterior chamber irrigation the vision improved to "rv 1.2 (~20/16), lv 1.0 (20/20)".Reportedly, bilateral tass was observed as well.In the reporter's opinion, the device did not fail to perform as intended and cause of the event is unknown.On day 4, the issue was reported as unresolved.The lens remains implanted.
 
Additional Manufacturer Narrative
H6: type of investigation 3331 - device history record review indicates that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key19068386
Report Number2023826-2024-01281
Device Sequence Number6901181
Product Code MTA
UDI-Device Identifier00840311302109
UDI-Public00840311302109
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberVICMO12.6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/04/2024
Initial Report FDA Received Date04/08/2024
Supplement Report FDA Received Date07/19/2024
Date Device Manufactured11/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL: SFC-45 - LOT# UNK; FOAM TIP PLUNGER - LOT# UNK; INJECTOR MODEL: MSI-PF - LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
Patient SexFemale
-
-