| Catalog Number |
FSENCOR01 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, the driver initially displayed an error code.It was further reported that the driver started to sample on its own and stopped after a while by it's self.Reportedly, the machine was restarted and the procedure was completed.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the serial number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a: through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the encor foot pedal was returned for evaluation.The encor foot pedal was visually inspected upon receipt and was found to be in good condition.The device was functionally tested and passed the tests during evaluation.No other anomalies were identified.Therefore, the investigation is determined to be unconfirmed for the reported device unintended movement issue.The root cause for reported unintended movement issue cannot be determined as the problem could not be reproduced.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.H10: g3.H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|
| |
|
Event or Problem Description
|
|
It was reported that during a biopsy procedure, the driver initially displayed an error code.It was further reported that the driver started to sample on its own and stopped after a while by it's self.Reportedly, the machine was restarted and the procedure was completed.There was no reported patient injury.
|
| |
|