Catalog Number FSENCOR01
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a biopsy procedure, the driver initially displayed an error code.It was further reported that the driver started to sample on its own and stopped after a while by it's self.Reportedly, the machine was restarted and the procedure was completed.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the serial number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a: through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: a device history record and manufacturing review was not required as the event was determined to be expected and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the encor foot pedal was returned for evaluation.The encor foot pedal was visually inspected upon receipt and was found to be in good condition.The device was functionally tested and passed the tests during evaluation.No other anomalies were identified.Therefore, the investigation is determined to be unconfirmed for the reported device unintended movement issue.The root cause for reported unintended movement issue cannot be determined as the problem could not be reproduced.Labeling review: the labeling/packaging review was not required as the reported event is not associated with a labeling, packaging, or use related issue.H10: g3.H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during a biopsy procedure, the driver initially displayed an error code.It was further reported that the driver started to sample on its own and stopped after a while by it's self.Reportedly, the machine was restarted and the procedure was completed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCOR FOOT PEDAL
Common Device Name
REUSABLE BIOPSY
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19068972
Report Number2020394-2024-00469
Device Sequence Number4657762
Product Code KNW
UDI-Device Identifier00801741086557
UDI-Public(01)00801741086557
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K013641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberFSENCOR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/11/2024
Supplement Date Received by Manufacturer04/12/2024
Initial Report FDA Received Date04/09/2024
Supplement Report FDA Received Date04/18/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-