Catalog Number 1816MSK
Medical Device Problem Codes Nonstandard Device (1420); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a computed tomographic guided lung biopsy procedure, the sharp inner metal stylet was allegedly detached from the plastic hub and the needle.It was further reported that a break was allegedy appeared to be where the metal stylet met the plastic hub.A coaxial was used and the procedure completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiration date: 04/2026).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one trocar stylet was received without the hub.During visual evaluation, it appeared the needle was detached from the hub, the needle appeared to have residue throughout and no other anomalies were observed.Using a black light (under sandblasting process), it was noted no adhesive residue was on the trocar stylet where the hub would have been.Upon dimensional observation, the length of the returned stylet measured and the measured within specification.Therefore, the investigation is confirmed for the reported detachment of device as the returned trocar stylet was detached from its hub and the investigation is confirmed for the identified nonstandard device as a lack of adhesive residue noted on the trocar stylet.The cause for the reported detachment of device or device component and the identified nonstandard device was traced to be a manufacturing related issue.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 04/2026), g3, h6 (device).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a computed tomographic guided lung biopsy procedure, the sharp inner metal stylet was allegedly detached from the plastic hub and the needle.A coaxial was used and the procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MISSION KIT
Common Device Name
BIOPSY INSTRUMENT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CAREFUSION D.R. 203 LTD.
zona franca las americas
km 22-e-1
santo domingo 11606
DR   11606
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key19069100
Report Number2020394-2024-00471
Device Sequence Number6891407
Product Code KNW
UDI-Device Identifier00801741097065
UDI-Public(01)00801741097065
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K171953
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 06/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number1816MSK
Device Lot Number0001516485
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/13/2024
Supplement Date Received by Manufacturer06/20/2024
Initial Report FDA Received Date04/09/2024
Supplement Report FDA Received Date06/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-