VERYAN MEDICAL LTD BIOMIMICS 3D VASCULAR STENT SYSTEM Back to Search Results
Catalog Number 142122-14
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Ischemia (1942); Pain (1994); Restenosis (4576)
Date of Event 11/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to interventions, ischemia, claudication/leg pain, and worsening of peripheral arterial disease leading to additional surgical intervention, endovascular intervention or amputation are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
 
Event or Problem Description
The patient was treated as part of the mimics 3d usa post-market observational study.On (b)(6) 2021, the patient was implanted with one biomimics 3d (bm3d) stent in the left leg, a 6.0 x 125 mm stent which was used to treat a denovo lesion of the right superficial femoral artery (sfa) distal third to proximal popliteal.An antegrade approach was used and the lesion was prepared using atherectomy.On (b)(6) 2023, the site identified a restenosis of treated segment (target lesion).It was reported as possibly related to the device and not related to the procedure.It was target lesion-related.It required a percutaneous intervention on (b)(6) 2023 as the subject had rutherford class (rc) 4 symptoms.The intervention was performed on the sfa distal third to anterior tibial distal third.It involved percutaneous transluminal angioplasty (pta)/ standard balloon angioplasty, laser atherectomy and thrombectomy.It was reported as a target lesion/vessel revascularisation (tlr/tvr).The patient outcome was reported as resolved/recovered.The device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Common Device Name
BIOMIMICS 3D VASCULAR STENT SYSTEM
Manufacturer (Section D)
VERYAN MEDICAL LTD
block 5
parkmore east business park
galway, H91 V 0TX
EI  H91 V0TX
Manufacturer Contact
alan mcdonagh
block 5
parkmore east business park
galway, H91 V-0TX
EI   H91 V0TX
MDR Report Key19069238
Report Number3011632150-2024-00018
Device Sequence Number4657779
Product Code NIP
UDI-Device Identifier05391526850480
UDI-Public(01)05391526850480(17)230109(11)210610(10)0000092737
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P180003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/09/2023
Device Catalogue Number142122-14
Device Lot Number0000092737
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2024
Initial Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN.; CLOPIDOGREL (PLAVIX).
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight71 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-