STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Catalog Number 625-0T-28D
Medical Device Problem Code Noise, Audible (3273)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 03/27/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that "patient allegedly felt some squeaking and mild grinding of her hip implant.She claims that her hip gave out while stepping out of the shower.Patient was being revised for possible fractured ceramic head.".
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.H3 other text : device not returned to the manufacturer.
 
Additional Manufacturer Narrative
Correction: investigation conclusion reveals that this reported device is not a suspected cause or contributor to the patient's experience and considered concomitant.An event regarding crack/fracture of a ceramic head involving a trident liner was reported.Conclusion: based on the information provided there is no indication or allegation that the device reported in this investigation contributed to the event and is therefore considered concomitant.Visual inspection of the returned devices indicated that the ceramic head had fractured into many pieces.The damage observed to the bearing surface of the returned trident liner is consistent with contact with the femoral component as well as debris from the fractured ceramic head.As the fracture of the ceramic head necessitated revision of the liner as well, the liner is considered concomitant.A full investigation is completed on the ceramic head within the same pi.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that "patient allegedly felt some squeaking and mild grinding of her hip implant.She claims that her hip gave out while stepping out of the shower.Patient was being revised for possible fractured ceramic head.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT ALUMINA INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19070220
Report Number0002249697-2024-00567
Device Sequence Number15166983
Product Code MRA
UDI-Device Identifier04546540516831
UDI-Public04546540516831
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P000013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2018
Device Catalogue Number625-0T-28D
Device Lot Number46089802
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2024
Supplement Date Received by Manufacturer08/10/2024
Initial Report FDA Received Date04/09/2024
Supplement Report FDA Received Date08/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age44 YR
Patient SexFemale
Patient Weight100 KG
-
-