| Model Number |
BWENO1-B |
| Medical Device Problem Code |
Use of Incorrect Control/Treatment Settings (1126)
|
| Health Effect - Clinical Code |
Blister (4537)
|
| Date of Event |
08/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Wife has been using her husband's unit for the past 1 1/2 to treat leg pain focused in her right hamstring.Never has had any issues unit yesterday.Later in the day after the treatment, she felt burning/itching and noticed a pus-filled blister near the area or possible under the area where one of the electrodes had been placed on her upper hamstring/glut area.No issue under the other pad which had been placed 6" apart on her lower hamstring.
|
| |
|
Additional Manufacturer Narrative
|
|
As the electrodes were not available to evaluate it is not possible to determine the root cause.However the most likely cause is that wire from one electrode might have pulled partially out of the electrodes contacting skin causing a blister.When a patient sits on the electrodes it is possible the leadwire cable can pull on the electrode wire.Another important point to be notes is that these electrodes were expired.
|
| |
|