| Model Number |
D284VRC |
| Medical Device Problem Codes |
Failure to Charge (1085); Electrical /Electronic Property Problem (1198); Delayed Charge Time (2586)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead triggered alerts for high and undefined high rv defibrillation coil impedance values.In addition, the rv lead triggered a lead integrity alert (lia) due to non-sustained ventricular tachycardia episodes and a high sensing integrity counter (sic) value.The lead also triggered an alert for out of range (oor) high rv pacing impedance value.The device reached end of service (eos) and triggered alerts for longer than expected charge time at elective replacement indicator (eri) and charge circuit timeout.Additionally, the device exhibited charge circuit inactive.The device was explanted and the lead is no longer in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|