MAUDE Adverse Event Report: TELEFLEX MEDICAL PU DP-52K DISP PUNCH 5.2MM; INSTRUMENTS, SURGICAL, CARDIOV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL PU DP-52K DISP PUNCH 5.2MM; INSTRUMENTS, SURGICAL, CARDIOV Back to Search Results
Model Number IPN915273
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Codes Perforation of Vessels (2135); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that "during use, punch gets jammed in the arteria".Additional information received states that "the punch tore the artery at the time of releasing and perforated the artery.The artery had to be cut to remove the punch.Arterial repair was required to prevent bleeding.The patient is fine".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Additional Manufacturer Narrative
(b)(4).One unit of catalog number dp-52k was received for analysis.Sample was not received in its original packaging.The sample has the tip release (no stuck) with signs of use.No other issues were observed.The punch was activated per work instruction, and no jammed issue were observed, the punch moved freely.A device history record review was performed, and no relevant findings were identified.Customer complaint cannot be confirmed, the unit had the tip release (no stuck) and no functional issues were observed during functional tests.However, complaints of this type will continue to be monitored via periodic reviews to evaluate if any trend exists.
 
Event or Problem Description
It was reported that "during use, punch gets jammed in the arteria".Additional information received states that "the punch tore the artery at the time of releasing and perforated the artery.The artery had to be cut to remove the punch.Arterial repair was required to prevent bleeding.The patient is fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PU DP-52K DISP PUNCH 5.2MM
Common Device Name
INSTRUMENTS, SURGICAL, CARDIOV
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
mariah mackinnon
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19071944
Report Number3004365956-2024-00026
Device Sequence Number15195238
Product Code DWS
UDI-Device Identifier14026704697974
UDI-Public14026704697974
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN915273
Device Catalogue NumberDP-52K
Device Lot Number74E2300016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2024
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer04/09/2024
Initial Report FDA Received Date04/09/2024
Supplement Report FDA Received Date04/23/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-