|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the tube's cuff presented two cuts.An inflation/deflation test was performed using a syringe; 16 cc of air was applied to the cuff, and it was observed the cuff deflated immediately.It was reported that the tracheostomy tube's cuff or pilot balloon had condensation during use.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ordinarily, cuff pressure should not exceed 25 cm of h2o.Over-inflation of the cuff may cause tracheal damage and may inhibit ventilation.Consult with the attending physician.Test each tube¿s cuff, pilot balloon, and valve by inflation prior to use.If dysfunction is detected in any part of the inflation system, the tube should not be used.Cuff inflation should be checked on a regular basis, and replacement tracheostomy tubes should be kept at bedside.To ease placement of the tracheostomy tube, generously lubricate the surface of the loading dilator.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, taper back the cuff by deflating the cuff and gently moving it away from the distal tip of the cannula toward the neck plate as the residual air is removed by deflation.When tapering the cuff, do not use sharp instruments that would damage the cuff, such as forceps or hemostats.Extremes of movement may result in mucosal abrasion and transient disruption of cuff seal.The use of lidocaine topical aerosol has been associated with the formation of pin holes in pvc cuffs.Lidocaine hydrochloride solution has been reported not to have this effect.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|