| Model Number |
0600-0783 |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer declined the option to have their glidescope core smart cable returned to verathon for evaluation.Since the device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history for the reported smart cable serial number "(b)(6)" did not identify any previous complaints.Trending analysis for glidescope core smart cables does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
A customer reported that during an emergency care procedure, using a glidescope core smart cable, the connected glidescope core monitor had no image and displayed a message indicating that no camera was connected even though a laryngoscope was connected.It was reported that this occurred in the middle of an intubation.The procedure was completed using a backup cable which was made available within a few minutes.No delay in the procedure or harm to the patient was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Following verathon's initial report submission, the customer returned the subject glidescope core smart cable to verathon for evaluation.A verathon technical service representative (tsr) evaluated the returned device and was able to confirm the reported cable failure.Upon visual inspection, the tsr identified two (2) bent pins inside the smart cable's hdmi connector.Due to the identified connector damage, the smart cable was unable to be connected to verathon's test equipment for evaluation of the reported image issue; however, it is likely that the damaged hdmi pins may have caused or contributed to the reported image issue.Upon completion of verathon's device evaluation, the customer's smart cable was scrapped due to there being no repairs available for the device.Corrective action is not required at this time.Verathon will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|