• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR FOGARTY ARTERIAL EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR FOGARTY ARTERIAL EMBOLECTOMY CATHETER; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 120804FP
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when opened this 120804fp fogarty catheter was found to have the balloon popped.There was no patient injury.Our product lab received one 120804f catheter.The reported issue of the balloon popped was confirmed.The balloon latex appeared deteriorated.There were multiple cracks and tears evident on the balloon latex.Leakage was observed through the tears on the balloon.Both balloon windings were intact.The balloon latex was released from proximal winding to check the balloon edges at the tears.The edges did not appear to match at the tears.Latex deterioration is a condition usually caused by age, excessive exposure to light, atmosphere, or ozone.Appears on the balloon surface as a maze of fine cracks or crazing.The condition may occur in a small area or cover the entire balloon.No other visible damage was observed from the catheter body.An engineering evaluation was initiated to assess for any manufacturing-related processes which could be correlated to the complaint.The storage conditions for these catheters are specified in the ifu.A product risk assessment addresses balloons with fragmentations for embolectomy catheters, and a corrective and preventive actions (capa) was created to address this issue.The capa investigation concluded that the root cause is exposure of latex to ozone, which can happen when the pouch packaged device is stored in rooms with high energy ionizing radiation sources that could generate ozone e.G.Fluoroscopy machines, x ray machines, uv lights, hvac sanitation equipment, etc.As a result, fca 90905 was voluntarily initiated instructing customers to return any product which has been stored in the same room as high energy ionizing radiation sources which emit ozone.A device history record review was completed and documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY ARTERIAL EMBOLECTOMY CATHETER
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
katheryn cespedes
1 edwards way
irvine, CA 92614
9492017706
MDR Report Key19075659
Report Number2015691-2024-02722
Device Sequence Number4798609
Product Code DXE
UDI-Device Identifier00690103205213
UDI-Public(01)00690103205213(17)241019(11)220721(10)64469758
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMEN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/19/2024
Device Model Number120804FP
Device Lot Number64469758
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2024
Initial Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
90905
Patient Sequence Number1
Patient SexUnknown
-
-