| Catalog Number |
46538BER/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The account alleges that while removing the device during the procedure, the catether detached within the patient.The clinician was able to successfully remove the detached piece without futher intervention.No additional patient consequence to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed, and the root cause is attributed to a weak fuse joint related to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|