• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CERENOVUS, INC. ENVOY 6F, MPD 100CM; PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CERENOVUS, INC. ENVOY 6F, MPD 100CM; PERCUTANEOUS CATHETER Back to Search Results
Catalog Number 67025800
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturer¿s ref.No:(b)(4).Information regarding patient identifier, weight, race, and ethnicity were not provided.Section e.1: the initial reporter phone: (b)(6).The initial reporter email address was not available / reported.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30914085) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.With the information available and without the complaint product available to be returned for analysis, the reported product issue documented in the complaint cannot be confirmed through functional evaluation and analysis.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint device; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Event or Problem Description
As reported via the health authority report, the envoy® guiding catheter, 6f, 100 cm, mpd (67025800 / 30914085) was inserted, it was observed fractured / cracked and the guide wire could not be delivered to the target site.The envoy catheter was replaced and the procedure was completed.There was no report of any negative patient impact.On 08-apr-2024, additional information was received.Per the information, the procedure was median cerebral imaging.The replacement device was another envoy® guiding catheter, 6f, 100 cm, mpd (67025800) the same guidewire was used after the guide catheter was replaced.The information confirmed there was no negative patient impact and no procedure delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVOY 6F, MPD 100CM
Common Device Name
PERCUTANEOUS CATHETER
Manufacturer (Section D)
CERENOVUS, INC.
6303 waterford district drive
suites 215 & 315
miami FL 33126
Manufacturer (Section G)
CERENOVUS INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key19076709
Report Number3007628272-2024-00021
Device Sequence Number20341068
Product Code DQY
UDI-Device Identifier10886704031734
UDI-Public10886704031734
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093184
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2025
Device Catalogue Number67025800
Device Lot Number30914085
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2024
Initial Report FDA Received Date04/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR