| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Intraocular Pressure Increased (1937); Blurred Vision (2137)
|
| Date of Event |
02/02/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.6mm vticm5_12.6; -13.0/1.5/121 (sphere/cylinder/axis) implantable collamer lens into the patient's left eye (os) on (b)(6)2024.The surgeon reports there is refractive surprise.The patient is on monoprost for iop.Lens remains implanted.The cause of the event was reported as unknown.The problem is not resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
It was later reported that on (b)(6) 2024 the lens was explanted and on this same date replaced with a same length but different power lens and the problem was resolved.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a micro-centrifuge vial, dry with residue/debris on the product.Visual inspection found no visible damage to the lens.· dimensional and functional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|