| Model Number |
HY10B07R5 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a custom tubing pack, it was reported that the extracorporeal membrane oxygenation (ecmo) circuit was plasmalyte primed and sat for almost 30 days with no leak.Once the customer then blood primed it for use, a leak started at the quick disconnect.The leak originated from the tubing connection.There was no patient blood loss as a result of the leak.An unplanned blood transfusion was not required.The same leak did not appear in multiple packs.The device was replaced.There was no patient involvement, so no adverse effect occurred.
|
| |
|