ZIMMER BIOMET, INC. NEXGEN STEMMED NONAUGMENTABLE TIBIAL COMPONENT CR/PS/LPS SIZE 5; PROSTHESIS, KNEE Back to Search Results
Catalog Number 00598604701
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Ambulation or Postural Difficulties (2544)
Date of Event 03/18/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient underwent an initial total knee arthroplasty.Approximately three years post-implantation, the patient presented with ongoing knee pain and difficulty walking due to a deep tissue infection.Subsequently, the patient underwent revision surgery where a loose tibial component was also revealed.Following revision surgery, the patient continued with ongoing antibiotic treatment for the infection.It was reported that no further information is available.
 
Additional Manufacturer Narrative
(b)(4).D10: medical product: femoral component option size f right compatible with lps-flex prolong: catalog#00596401652, lot#63784946; articular surface size ef 10 mm height: catalog#00596404010, lot#65307716.G2: foreign: united kingdom.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Upon receipt of additional information or completion of the investigation, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; g3; h2; h3; h6; h11.Visual and dimensional evaluations of the product could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: b4; b5; d9; g3; h2; h3; h6; h11.Visual evaluation of the returned products found evidence of use (surgical debris and surface scratches); however, no further information can be determined for the tibial component.Device was measured in areas accessible with no bone cement residue and found to be conforming to manufacturing specifications.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Reported event was unable to be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information at time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEXGEN STEMMED NONAUGMENTABLE TIBIAL COMPONENT CR/PS/LPS SIZE 5
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19080222
Report Number0001822565-2024-01205
Device Sequence Number5859641
Product Code JWH
UDI-Device Identifier00889024218857
UDI-Public(01)00889024218857(17)310330(10)J6994572
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number00598604701
Device Lot NumberJ6994572
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2024
Supplement Date Received by Manufacturer09/18/2024
01/15/2025
Initial Report FDA Received Date04/10/2024
Supplement Report FDA Received Date09/19/2024
02/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-