Catalog Number 04.211.036
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2: additional procode: hrs.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: initial reporter is a synthes employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from japan reports an event as follows: it was reported that on (b)(6) 2024, the patient underwent a transcondylar fracture of the distal humerus.After implanting a posterior lateral plate, the surgeon began implanting the medial plate, but three screws idled and no torque could be applied.The first screw was inserted after drilling using a 2.7 va drill guide, but the screw idled when the head came near to the plate.This screw was removed because a metal fragment was generated.The second screw was inserted into the same hole as the first, but it also idled.Torque could be lightly applied to this screw so it was left in the hole.The third screw was inserted into another hole after drilling with the va drill guide, but no torque could be applied.This screw was also left in the hole.Another drill guide was used and the issues with screw idling and lack of torque were resolved.The procedure was completed successfully with no surgical delay.There was no adverse patient impact, it was confirmed via x-ray that no fragments remained within the patient.No further information is available.This report is for a 2.7mm ti va lckng scr slf-tpng with t8 stardrive recess 36mm.This is report 4 of 5 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: part # 04.211.036s lot # 146l211 manufacturing site: werk selzach logistik release to warehouse date: 06 mar 2023 expiration date: 01 mar 2033 supplier: (b)(4).A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.H6 component code: g07002 (appropriate term/code not available) used to capture no findings available due to no product returned.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM TI VA LCKNG SCR SLF-TPNG WITH T8 STARDRIVE RECESS 36MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK SELZACH LOGISTIK
bohnackerweg 5
selzach 2545
SZ   2545
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19080485
Report Number8030965-2024-04760
Device Sequence Number6029956
Product Code HWC
UDI-Device Identifier10886982119100
UDI-Public(01)10886982119100
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.211.036
Device Lot Number146L211
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/26/2024
Supplement Date Received by Manufacturer04/16/2024
Initial Report FDA Received Date04/10/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VA DRILL GUIDE 2.7; VA LOCKSCR Ø2.7 HEAD 2.4 SELF-TAP L30 TA; VA LOCKSCR Ø2.7 HEAD 2.4 SELF-TAP L30 TA; VA-LCP DHP 2.7/3.5 MED R SHORT 1-HO L69
Patient SexUnknown
-
-