• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FREE PSA; PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FREE PSA; PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS Back to Search Results
Catalog Number 08828610190
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The analyzer serial number is (b)(6).The calibration and qc recovery data provided was acceptable.The patient samples were requested for investigation.The investigation is ongoing.
 
Event or Problem Description
There was an allegation of questionable elecsys free psa immunoassay results compared to total psa results for 1 patient on a cobas e 801 analytical unit compared to a siemens analyzer.This medwatch will cover free psa.Refer to medwatch with a1 patient identifier pt (b)(6) for information on the total psa results.On (b)(6) 2024, the patient had a sample tested on module 3 free psa result of 1.23 ng/ml and a total psa result of 0.101 ng/ml.On (b)(6) 2024, the sample was repeated on a siemens analyzer and the free psa result was 0.03 ng/ml and the total psa result was 1.04 ng/ml.On (b)(6) 2024, the patient had a sample tested on module 4 and the free psa result was 1.20 ng/ml and a total psa result of 0.103 ng/ml.On (b)(6) 2024, the sample was repeated on a siemens analyzer and the free psa result was 0.03 ng/ml and the total psa result was 1.00 ng/ml.
 
Additional Manufacturer Narrative
The investigation did not identify a product problem.The specific cause of the event could not be determined.Based on the provided data, the issue was consistent with the presence of rare endogenous interfering substance(s) in the patient samples.Per product labeling: it is known that in rare cases psa isoforms do exist which may be measured differently by different psa tests.Findings of this kind have occasionally been reported for psa tests from various manufacturers.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FREE PSA
Common Device Name
PROSTATE-SPECIFIC ANTIGEN (PSA) FOR MANAGEMENT OF PROSTATE CANCERS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key19080577
Report Number1823260-2024-01088
Device Sequence Number6030416
Product Code LTJ
UDI-Device Identifier07613336166628
UDI-Public07613336166628
Combination Product (Y/N)Y
Initial Reporter CountryEG
PMA/510(K) Number
P000027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Catalogue Number08828610190
Device Lot Number73065701
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/19/2024
Supplement Date Received by Manufacturer04/23/2024
Initial Report FDA Received Date04/10/2024
Supplement Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-