| Catalog Number |
368774 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while using the bd vacutainer® rapid serum tube (rst) thrombin, one tube had poor serum separation and re-mixed.No patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
E.1: initial reporter addr 1: 815 springfield commons dr (b)(6).The manufacturing location for this product is sekisui.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1., and the franklin lakes fda registration number has been used for the manufacture report number.H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using the bd vacutainer® rapid serum tube (rst) thrombin, one tube had poor serum separation and re-mixed.No patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation summary: material #: 368774, lot/batch #: 231011.Bd had not received samples, but 1 photo was provided for investigation.The photo was reviewed and a normal gel barrier separation was not observed.Additionally, retention samples from bd inventory were evaluated by functional testing for gel movement and thrombin strength, centrifuged at 1300g for 5 minutes at 15 degrees centigrade, and no issues were observed relating to poor barrier separation as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode poor barrier separation.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|