|
H3 visual findings observed one pair of restraints received for evaluation in the original packaging with ifus and a posey key.The restraints appear new or have not been put into use yet.No abnormalities have been observed.The reported issue was not confirmed.The buckles locked properly when in use with their respective straps.The cuff and bed connecting straps are found out to be within the manufacturing specifications.No abnormalities have been observed.Instructions for use (ifu) were reviewed and found to provide adequate instructions and warnings for safe and effective use of the device.The ifu states to before each use, check cuff s and straps for cracks, tears, and/or excessive wear or stretch, broken buckles or locks, and/or that hook-and-loop adheres securely as these may allow patient to remove cuff.Discard if device is damaged or if unable to lock.The ifu also provided extra warning to not use this device with someone who has continued highly aggressive or combative behavior, self-destructive behavior, or deemed to be an immediate risk to others or to self.At this time there is no evidence that a manufacturing non-conformity contributed to the reported complaint.Therefore, no corrective or preventative actions are necessary.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
|