MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP XLPE INSERT 43/22 NON-CEM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP XLPE INSERT 43/22 NON-CEM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED Back to Search Results
Catalog Number 75018942
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal reference number: (b)(4) h10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
It was reported that during thr surgery one (1) polarcup xlpe insert 43/22 non-cem was placed in the press with the 22 head.It seemed to have created a vacuum, as the head moved inside the insert and appeared to be fine.However, when placing the insert on the patient and reducing it, the insert came loose from the head.They had to remove the head from the stem.The process of inserting the head into the insert was repeated, and when reducing it again, the exact same issue occurred.The treatment was resumed after a significant delay, using a s+n back-up device, which was implanted correctly.No injuries were reported as a consequence of this issue.
 
Additional Manufacturer Narrative
It was reported that during total hip replacement surgery one (1) polarcup xlpe insert 43/22 non-cem was placed in the press with the 22 head.It seemed to have created a vacuum, as the head moved inside the insert and appeared to be fine.However, when placing the insert on the patient and reducing it, the insert came loose from the head.They had to remove the head from the stem.The process of inserting the head into the insert was repeated, and when reducing it again, the exact same issue occurred.The treatment was resumed after a significant delay, using a s+n back-up device, which was implanted correctly.The device intended for use in treatment was not returned for investigation.A product evaluation was not possible.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The review of historical complaints for the alleged device revealed no additional similar complaints reported for the same batch, and no additional similar complaints for the same product number over the past 12 months with similar failure mode.Review of past corrective actions was performed.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation does not lead to an accurately determined cause.There is no indication that the reported devices failed to meet manufacturing specifications upon release for distribution.There is no need for further actions.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP XLPE INSERT 43/22 NON-CEM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL/POLYMER, CEMENTED OR UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19084373
Report Number9613369-2024-00036
Device Sequence Number15524735
Product Code KWY
UDI-Device Identifier07611996111323
UDI-Public07611996111323
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K110135
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75018942
Device Lot NumberB2315699
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2024
Supplement Date Received by Manufacturer05/01/2024
Initial Report FDA Received Date04/10/2024
Supplement Report FDA Received Date05/08/2024
Date Device Manufactured08/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-