• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404282
Medical Device Problem Codes Collapse (1099); Deflation Problem (1149); Inflation Problem (1310); Defective Device (2588); Material Deformation (2976); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 04/01/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced discomfort.An attempt was made to activate the device; however, it was found that the device does not inflate nor deflate being a blockage in the tube between the reservoir and pump the suspected cause.In addition, the pump was reported to stay flat when pressed.A forced deflation maneuver was performed.The device remains implanted, and a revision surgery has not been scheduled.No additional patient complications were reported.
 
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced discomfort.An attempt was made to activate the device; however, it was found that the device does not inflate nor deflate being a blockage in the tube between the reservoir and pump the suspected cause.In addition, the pump was reported to stay flat when pressed.A forced deflation maneuver was performed.Additional information reported that on july 5, 2024, the explant surgery was performed.It was noted that the tube from the pump to the reservoir was twisted many times on both sides.The device was removed with no additional patient complications reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, this device was thoroughly analyzed.The first cylinder had wear at a fold in the middle and distal end.The first cylinder passed leakage testing.The second cylinder had a hole with a leak in the proximal end from sharp instrument damage.The momentary squeeze (ms) pump passed visual inspection, leakage testing and functional testing.The flat reservoir passed visual inspection and leakage testing.
 
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced discomfort.An attempt was made to activate the device; however, it was found that the device does not inflate nor deflate being a blockage in the tube between the reservoir and pump the suspected cause.In addition, the pump was reported to stay flat when pressed.A forced deflation maneuver was performed.Additional information reported that on (b)(6) 2024, the explant surgery was performed.It was noted that the tube from the pump to the reservoir was twisted many times on both sides.The device was removed with no additional patient complications reported.
 
Event or Problem Description
It was reported that the patient with this inflatable penile prosthesis (ipp) experienced discomfort.An attempt was made to activate the device; however, it was found that the device does not inflate nor deflate being a blockage in the tube between the reservoir and pump the suspected cause.In addition, the pump was reported to stay flat when pressed.A forced deflation maneuver was performed.The device remains implanted, and a revision surgery has not been scheduled.No additional patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key19084557
Report Number2124215-2024-20669
Device Sequence Number8359212
Product Code FHW
UDI-Device Identifier00878953003856
UDI-Public00878953003856
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/14/2026
Device Model Number72404282
Device Catalogue Number72404282
Device Lot Number1000529485
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/01/2024
Supplement Date Received by Manufacturer05/01/2024
07/05/2024
09/03/2024
Initial Report FDA Received Date04/10/2024
Supplement Report FDA Received Date05/16/2024
07/16/2024
09/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RESERVOIR: UPN: 720182-01 / LOT:1100198082.; RESERVOIR: UPN: 720182-01 / LOT:1100198082.; RESERVOIR: UPN: 720182-01 / LOT:1100198082.; RESERVOIR: UPN: 720182-01 / LOT:1100198082.
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient Age56 YR
Patient SexMale
-
-