| Catalog Number |
1588610122 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: surgery issue: picture is bad.Update: blurry the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause could be the camera head was pulled from the cable and improper cleaning sterilization.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image.
|
| |
|