| Model Number |
26006M |
| Medical Device Problem Codes |
Fire (1245); Flare or Flash (2942)
|
| Health Effect - Clinical Codes |
Bowel Burn (1756); Burn(s) (1757)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts were made to obtain more information and to have reported device return for evaluation; however, the reporter has not responded to the requests at the time of this report.If the reported device will be returned or the evaluation is completed, a supplemental report would be made to the fda.This complaint will be tracked and trended.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the unipolar high frequency cord sparked and caught fire during the surgery.While using the electrode to cauterize, the center area of the cord sparked and caught surgical drape and arm cradle foam padding on fire.A small amount of water was used to extinguish the fire.The patient sustained a burn on he inner left forearm and also several spots on the left side of abdomen.
|
| |
|
Additional Manufacturer Narrative
|
|
Changes made from the initial report: updated section d9 to no.Per investigation by the manufacturing site: it was reported that the unipolar high frequency cord sparked and caught fire during the surgery.While using the electrode to cauterize, the center area of the cord sparked and caught surgical drape and arm cradle foam padding on fire.The patient sustained a burn ont he inner left forearm and also several spots on the left side of abdomen.Most probable root cause: this issue occurs if the cable is used over its intended lifetime.Due to external influences/forces, the electrical resistance increases, and the cable becomes hot or burns.According to the ifu, this issue could happen in given circumstances.Based on the given facts we set the root cause to wearing.This event is filed under internal complaint id (b)(4).
|
| |
|
Search Alerts/Recalls
|