Model Number 757
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer sent a bur in with the device.The bur used had a shaft length that was not recommended by manufacturer.This may have contributed to the bur coming out.No manufacturing or material defects were noted.
 
Event or Problem Description
Customer reported the bur came out during a procedure.The bur was collected by doctor without injuring the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSTYLE SF
Common Device Name
HIGH-SPEED DENTAL HANDPIECE
Manufacturer (Section D)
LARES RESEARCH
295 lockheed avenue
chico CA 95973
Manufacturer (Section G)
LARES RESEARCH
295 lockheed avenue
chico CA 95973
Manufacturer Contact
bruce holderbein
295 lockheed avenue
chico, CA 95973
5303451767
MDR Report Key19085342
Report Number2916440-2024-00004
Device Sequence Number18067094
Product Code EFB
UDI-Device IdentifierD950136210360
UDI-PublicD950136210360
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K141221
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number757
Device Catalogue Number13621-036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2024
Initial Report FDA Received Date04/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/29/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-