| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was evaluated by zoll medical corporation.The customer's report was duplicated and attributed to a defective front enclosure.The front enclosure was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of the report of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|