• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC SYMPHONY OCT SYSTEM REDUCER TOWER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC SYMPHONY OCT SYSTEM REDUCER TOWER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 202000141
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.B3: unknown what date device broke.E1: initial reporter is j&j company representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported during a training session the kit was delivered broken and never used for procedures.It is unknown when the kit broke.No patient involvement.This report is for one symphony oct system reducer tower for (b)(4).
 
Event or Problem Description
It was reported that the prongs on the inner tubes are squished together, meaning rod reduction cannot occur.Also, there is no capture on the inner tube meaning it constantly slides out of the tower, therefore these instruments cannot be used for operations.The issue was discovered during a training session.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMPHONY OCT SYSTEM REDUCER TOWER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key19086614
Report Number1526439-2024-01278
Device Sequence Number20726931
Product Code LXH
UDI-Device Identifier10705034530574
UDI-Public(01)10705034530574
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number202000141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer04/15/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date05/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-