| Model Number |
CH1030 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The henly retractor set subject of the reported was returned for evaluation.Investigation of this event is currently in process.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that during a patient procedure, the blades to their henly retractor were not "locking on" and "sliding off" during use.User facility personnel stated the patient procedure was completed successfully and no injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The henly retractor set subject of the reported was returned for evaluation.Evaluation of the device determined that no issues were noted with the function or operation.The device operated according to specifications and the reported event could not be duplicated.The device history record (dhr) was reviewed and confirmed the device was manufactured to specification; no abnormalities were noted.A complaint review was performed and confirmed this event to be isolated for the lot subject of the reported event.No additional issues have been reported.
|
| |
|