| Model Number |
EEAXL25 |
| Medical Device Problem Codes |
Break (1069); Failure to Form Staple (2579); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); Tissue Breakdown (2681)
|
| Date of Event |
03/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during an ivor lewis esophagectomy procedure, when performing the anastomosis, the surgeon was squeezing the handle stating that the safety was not letting go.The surgeon kept squeezing, a snap was heard and the safety lever broke.All pieces of the lever was found.The surgeon kept using the stapler, when the handle was squeezed, the usual crunch sound was not heard.When the stapler was pulled the anvil was still attached, and the tissue rings were intact, but the ring part of the stapler was missing.Further dissection was required to do another anastomosis, and while doing it, they found inside the patient a full ring of staples loose as the staples did not form properly and the ring part of the stapler where it attaches to the anvil.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but three photos were available for evaluation.Visual inspection noted the safety and staple guide are disengaged and the staple guide appears to have a full complement of staples.It was reported that a component disengaged, handle broken, instrument made strange noise, no audible feedback, and staple formation.The reported issues were confirmed.The most likely cause could not be established from the information available.This issue may occur if the latch is forced into disengagement prior to green indicator visibility.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the safety will not release if the green bar is not visible in the indicator window.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|