US SURGICAL PUERTO RICO EEA; STAPLER, SURGICAL Back to Search Results
Model Number EEAXL25
Medical Device Problem Codes Break (1069); Failure to Form Staple (2579); Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949); Noise, Audible (3273)
Health Effect - Clinical Codes Failure to Anastomose (1028); Tissue Breakdown (2681)
Date of Event 03/14/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during an ivor lewis esophagectomy procedure, when performing the anastomosis, the surgeon was squeezing the handle stating that the safety was not letting go.The surgeon kept squeezing, a snap was heard and the safety lever broke.All pieces of the lever was found.The surgeon kept using the stapler, when the handle was squeezed, the usual crunch sound was not heard.When the stapler was pulled the anvil was still attached, and the tissue rings were intact, but the ring part of the stapler was missing.Further dissection was required to do another anastomosis, and while doing it, they found inside the patient a full ring of staples loose as the staples did not form properly and the ring part of the stapler where it attaches to the anvil.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but three photos were available for evaluation.Visual inspection noted the safety and staple guide are disengaged and the staple guide appears to have a full complement of staples.It was reported that a component disengaged, handle broken, instrument made strange noise, no audible feedback, and staple formation.The reported issues were confirmed.The most likely cause could not be established from the information available.This issue may occur if the latch is forced into disengagement prior to green indicator visibility.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the safety will not release if the green bar is not visible in the indicator window.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLER, SURGICAL
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19088460
Report Number2647580-2024-01583
Device Sequence Number18032102
Product Code GAG
UDI-Device Identifier10884523006056
UDI-Public10884523006056
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K221003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEEAXL25
Device Catalogue NumberEEAXL25
Device Lot NumberP3D0164
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2024
Supplement Date Received by Manufacturer05/31/2024
Initial Report FDA Received Date04/11/2024
Supplement Report FDA Received Date06/19/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-